Philips North America Llc recalls Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator designed fo
Recalled November 22, 2023 · FDA recall Z-0294-2024 · Philips North America
Description
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.