Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini GXL 16 Slice System Number: 882410 Continu

Recalled December 1, 2021 · FDA recall Z-0278-2022 · Philips North America

Hazard
Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part · FDA Class II
Units
1 unit
Sold at
US (nationwide) distribution to states of: CA, ME, TX and OUS (Foreign) to countries of: Australia, Germany, Netherlands and Saudi Arabia.

Description

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

Product
Continuous CT Fluoroscopy DVI Ceiling Mount - CCT Arm DVI Option - Gemini GXL 16 Slice System Number: 882410 Continuous CT (CCT) is a CT or PET/CT product option that enables physicians and radiologists to produce real-time CT images to facilitate image-guided surgery.
Manufacturer(s)
Philips North America Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov