Philips North America Llc recalls Evolution upgrade 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnosti
Recalled May 29, 2024 · FDA recall Z-1843-2024 · Philips North America
Hazard
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire · FDA Class II
Units
56 units
Sold at
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Description
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Product
Evolution upgrade 1.5T- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): (1) 782116; (2) 782148
Manufacturer(s)
Philips North America Llc