Philips North America Llc recalls Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109
Recalled June 16, 2021 · FDA recall Z-1795-2021 · Philips North America
Hazard
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet. · FDA Class II
Units
156 MR Systems
Sold at
Domestic: FL, GA, MI, MO, NH, NJ, NY, PA, SC, TX, VT. Foreign: Australia, Austria, Bangladesh, Belgium, Brazil, Denmark, Finland, France, Germany, Iraq, Italy, Japan, Libya, Moldova, Netherlands, Oman, Spain, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Description
The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
Product
Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109 - Ingenia Ambition X
Manufacturer(s)
Philips North America Llc