Philips North America Llc recalls Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can

Recalled September 1, 2021 · FDA recall Z-2345-2021 · Philips North America

Hazard
Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury · FDA Class II
Units
11 units
Sold at
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.

Description

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Product
Ingenia CX 3.0T-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 78127 (OUS)
Manufacturer(s)
Philips North America Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov