Philips North America, LLC recalls IntelliVue MX800 patient monitor, model no. 865240
Recalled July 29, 2020 · FDA recall Z-2696-2020 · Philips North America
Hazard
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on. · FDA Class II
Units
28
Sold at
Domestic distribution to CA, HI, MA, NY, and UT. International distribution to Costa Rica, Ireland, Lebanon, and Poland.
Description
Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.
Product
IntelliVue MX800 patient monitor, model no. 865240
Manufacturer(s)
Philips North America, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →