RecallWatchUSA

Home / Recalls / Medical Devices

Philips North America, LLC recalls MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for

Recalled February 14, 2018 · FDA recall Z-0489-2018 · Philips North America

Hazard
Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode. · FDA Class I
Units
71 units
Sold at
No U.S. distribution, Internationally to: India, Ireland, Japan, Republic of Korea, Malaysia, Mexico, South Africa & United Kingdom

Description

Affected units may have a defective component that is intended to protect the internal circuitry from high voltages during shock delivery. The defective component may also prevent delivery of shocks in AED mode.

Product
MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Manufacturer(s)
Philips North America, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips North America, LLC recalls MRx Defibrillator Model # M3535A & M3536A, UDI 00884838000018(21)US00601969 Product Usage: The HeartStart MRx is for | RecallWatchUSA