Philips North America Llc recalls Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT
Recalled October 1, 2025 · FDA recall Z-2595-2025 · Philips North America
Hazard
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement. · FDA Class II
Units
181 units
Sold at
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.
Description
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Product
Philips CT systems labeled as the following: 1. Brilliance CT 40 Channel, Model Number: 728235. 2. Brilliance CT 64 Channel, Model Number: 728231. 3. Ingenuity Flex, Model Number: 728317
Manufacturer(s)
Philips North America Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →