Philips North America, LLC recalls Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M

Recalled July 17, 2019 · FDA recall Z-1923-2019 · Philips North America

Hazard
The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery. · FDA Class II
Units
97,794 Units
Sold at
Nationwide domestic distribution. Worldwide distribution.

Description

The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery.

Product
Philips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9
Manufacturer(s)
Philips North America, LLC