Philips North America Llc recalls Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radio
Recalled May 10, 2023 · FDA recall Z-1477-2023 · Philips North America
Hazard
Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking · FDA Class II
Units
Total Quantity: 234 units. US 195units. R.O.W 39 units
Sold at
Worldwide distribution - US Nationwide including Puerto Rico and the countries of Australia, Chile, China, Germany, India, Italy, Malaysia, Netherlands, Panama, Spain, Thailand.
Description
Potential table stop due to a broken tabletop cable: On the ProxiDiagnost N90 table there is a cable located under the tabletop that is covered by a plastic sleeve which can crack and damage the cable. If the tabletop cable breaks, the user may not be able to initiate table tilting movements or braking
Product
Philips ProxiDiagnost N90 1.0 (Model 706100) and ProxiDiagnost N90 1.1 (Model 706110)- A multi-functional general radiography and fluoroscopy (R/F) system
Manufacturer(s)
Philips North America Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →