Philips North America Llc recalls Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

Recalled December 25, 2024 · FDA recall Z-0686-2025 · Philips North America

Hazard
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data. · FDA Class II
Units
4,811 units
Sold at
Nationwide distribution

Description

The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.

Product
Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Manufacturer(s)
Philips North America Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips North America Llc recalls Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216 | RecallWatchUSA