Philips North America Llc recalls Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5

Recalled July 1, 2026 · FDA recall Z-2579-2026 · Philips North America

Hazard
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system. · FDA Class II
Units
909 units
Sold at
US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.

Description

Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.

Product
Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
Manufacturer(s)
Philips North America Llc