RecallWatchUSA

Home / Recalls / Medical Devices

Philips North America, LLC recalls Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minima

Recalled May 6, 2020 · FDA recall Z-1801-2020 · Philips North America

Hazard
No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose.. · FDA Class II
Units
162 units
Sold at
Worldwide distribution: US (nationwide) including the following states: Arkansas, Arizona, California, Colorado, Delaware, Florida, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin. And the following Countries of: Australia, Azerbaijan, Chile, Croatia, Czechoslovakia, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Korea South, Lebanon, Mexico, Netherlands, Norway, Pakistan, Poland, Russian Federate, Senegal, Spain, South Africa, Sweden, Switzerland, Taiwan, and Uruguay.

Description

No torque was specified for the four screws that connect the gearbox and the mounting flange within the C-arm roll motor which could result in the screws becoming loose..

Product
Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minimally invasive surgery procedures for the following clinical application areas: vascular, non-vascular, cardiovascular and neuro procedures.
Manufacturer(s)
Philips North America, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips North America, LLC recalls Phillips Azurion 7 M20, Angiographic X-Ray System Code 722079 Image guidance in diagnostic, interventional and minima | RecallWatchUSA