Philips North America Llc recalls SmartPath to dStream for 3.0T Model Number (REF): 782145
Recalled May 29, 2024 · FDA recall Z-1861-2024 · Philips North America
Hazard
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire · FDA Class II
Units
3 units
Sold at
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Description
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Product
SmartPath to dStream for 3.0T Model Number (REF): 782145
Manufacturer(s)
Philips North America Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →