Hazard
The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms. · FDA Class II
Units
21,672
Sold at
The products were distributed US nationwide. Global distribution.

Description

The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.

Product
TRx4851A 2.4 GHz IntelliVue Tele TRX, Model Number 862231; Includes Service Numbers 453564052401, 453564052411, 453564052441, 453564052451, 453564166851, 453564166861
Manufacturer(s)
Philips North America, LLC