Home / Recalls / Medical Devices
Philips North America, LLC recalls VesselNavigator application used with Philips Azurion (Azurion 7 Series) Software R1.2 and R2.0, and Philips Allura Xper
Recalled February 19, 2020 · FDA recall Z-1145-2020 · Philips North America
Hazard
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction. · FDA Class II
Units
443
Sold at
Worldwide Distribution. US nationwide, Canada, Argentina, Australia, Austria, Belgium, Brazil, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Republic of Korea, Latvia, Libya, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Philippines, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Uruguay.
Description
Due to a software defect, when a digital subtraction angiography (DSA) is exported to the VesselNavigator application, the DSA is displayed without subtraction.
Product
VesselNavigator application used with Philips Azurion (Azurion 7 Series) Software R1.2 and R2.0, and Philips Allura Xper R9, upgraded to Azurion 1.2 M20. System codes: Azurion 7 B12 (722067), Azurion 3 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079), Field Extensions Azurion 7 series (722136), Allura Xper R9 (722303). The product is software, used during an endovascular intervention procedure.
Manufacturer(s)
Philips North America, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →