RecallWatchUSA

Home / Recalls / Medical Devices

Philips North America recalls Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

Recalled November 22, 2023 · FDA recall Z-0306-2024 · Philips North America

Hazard
When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted. · FDA Class II
Units
1 unit
Sold at
US Nationwide distribution in the state of MN.

Description

When preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.

Product
Spectral CT on Rails, Software Version 5.1.0, Model Number 728334
Manufacturer(s)
Philips North America

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov