RecallWatchUSA

Home / Recalls / Medical Devices

Philips Respironics, Inc. recalls BiPAP V30 Auto Ventilator, Part Number 1111178

Recalled May 29, 2024 · FDA recall Z-1812-2024 · Philips Respironics

Hazard
Potential for devices to experience interruption/loss of therapy during a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients. · FDA Class I
Units
5,958 US; None OUS
Sold at
Worldwide distribution. US Nationwide including PR and GU; Austria, Belgium, Cyprus, Czech Republic, Denmark, France, Germany, Hungary, Italy, Netherlands, Poland, Portugal, Romania, Serbia, Slovakia, Spain, and Turkey.

Description

Potential for devices to experience interruption/loss of therapy during a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.

Product
BiPAP V30 Auto Ventilator, Part Number 1111178
Manufacturer(s)
Philips Respironics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Respironics, Inc. recalls BiPAP V30 Auto Ventilator, Part Number 1111178 | RecallWatchUSA