Home / Recalls / Medical Devices
Philips Respironics, Inc. recalls Garbin EVO Ventilator, Linde - Model Number: LD2110X23B Trilogy Evo is intended for pediatric through adult patients
Recalled February 26, 2020 · FDA recall Z-0946-2020 · Philips Respironics
Hazard
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439. · FDA Class I
Units
105
Sold at
US: None OUS: CH DE DK IT FR IE GB KW QA PL ES AT PH AE SI CZ
Description
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.
Product
Garbin EVO Ventilator, Linde - Model Number: LD2110X23B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.
Manufacturer(s)
Philips Respironics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →