RecallWatchUSA

Home / Recalls / Medical Devices

Philips Respironics, Inc. recalls Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B

Recalled January 10, 2024 · FDA recall Z-0638-2024 · Philips Respironics

Hazard
The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2. · FDA Class II
Units
100,126 (54,691 US; 45,435 OUS) in total
Sold at
Worldwide distribution: US (nationwide)-AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, Including PR; and OUS (International) including countries of: AE, AR,AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, DZ, ;EC EE, EG, ES, FI, FR, GB, GE, GF, GR, GT, HK, HR, HU, IE, IL, IN, IT, JO, JP, KE, KR, KW, LB, LT, MA, ME, MK, MM, MQ, MT, MX, MY, NG, NL, NO, NP, OM, PA, PE, PF, PH, PK, PL, PR, PT, PY, QA, RE, RO, RS, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW, UA, YT, ZA and ZW.

Description

The following was missing from the Contraindications Statement: The AVAPS-AE therapy mode is contraindicated for invasive use and patients less than 10 kg. The AVAPS feature is contraindicated for patients less than 10 kg. Potential harms if devices are used on contraindicated patients include barotrauma, hypoventilation/hypercapnia, and rebreathing of excessive CO2.

Product
Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B DS2110X11B EE2100X15B EE2110X15B ES2100X15B ES2110X15B EU2100X15B EU2100X19 EU2110X15B FR2100X14B FR2110X14B FX2100X15B GB2110X15B IA2100X15B IA2110X15B IN2100X15B IN2100X19 IN2110X15B IT2100X21B IT2110X21B JP2100X16B JP2110X16B KR2110X15B LA2100X15B LA2110X15B ND2110X15B RDE2100X13B RDE2110X13B RDS2110X11B RIN2100X15B RIN2110X15B RJP2100X16B RJP2110X16B UDS2110X11B
Manufacturer(s)
Philips Respironics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Respironics, Inc. recalls Trilogy Evo and Trilogy Evo O2 Material Numbers BL2110X15B BR2110X18B CA2110X12B DE2100X13B DE2110X13B DS2100X11B | RecallWatchUSA