RecallWatchUSA

Home / Recalls / Medical Devices

Philips Respironics, Inc. recalls Trilogy Evo O2 and LifeVentEVO2 Portable Electric Ventilator, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15

Recalled April 12, 2023 · FDA recall Z-1287-2023 · Philips Respironics

Hazard
The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen. · FDA Class I
Units
62,826 total devices
Sold at
Worldwide distribution. US Nationwide including Puerto Rico, and countries of AE, AR, AT, AU, BD, BE, BG, BO, BR, CA, CH, CL, CN, CO, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GE, GT, HK, HR, HU, ID, IL, IE, IN, IT, JP, KW, LB, LT, MM, MQ, MX, MY, NG, NL, NO, NP, NZ, PE, PF, PH, PK, PL, PT, QA, RO, RS, RU, SA, SG, SI, SK, TH, TR, TW, UA, VN, and ZA.

Description

The accuracy of delivered oxygen may deviate below the required tolerance of 5% from setpoint when providing high concentration oxygen therapy. Additionally, if equipped, the internal FiO2 sensor may indicate a value higher than the device is actually delivering. This may lead to under delivery of oxygen.

Product
Trilogy Evo O2 and LifeVentEVO2 Portable Electric Ventilator, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15B, JP2100X16B, AU2100X15B, LA2100X15B, CA2100X12B, CN2100X17B, BR2100X18B, KR2100X15B, IA2100X15B, PP2100X10, FP2100X10, FR2100X14B, ND2100X15B, IT2100X21B, ES2100X15B, DE2100X13B, BL2100X15B, GB2100X15B, EU2100X15B, EU2100X19, EE2100X15B, TR2100X15B, UDS2100X11B, RDS2100X11B, and SP2100X26B
Manufacturer(s)
Philips Respironics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Respironics, Inc. recalls Trilogy Evo O2 and LifeVentEVO2 Portable Electric Ventilator, Model Numbers DS2100X11B, IN2100X15B, IN2100X19, FX2100X15 | RecallWatchUSA