RecallWatchUSA

Home / Recalls / Medical Devices

Philips Respironics, Inc. recalls Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X

Recalled August 25, 2021 · FDA recall Z-2287-2021 · Philips Respironics

Hazard
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage). · FDA Class II
Units
18588 (US); 5662 (OUS)
Sold at
Worldwide distribution - US nationwide distribution.

Description

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

Product
Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X19, FR2110X14B, GB2110X15B, IA2110X15B, IN2110X15B, IN2110X19, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B, ND2110X15B, RDE2110X13B, UDS2110X11B
Manufacturer(s)
Philips Respironics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Respironics, Inc. recalls Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X | RecallWatchUSA