Philips Respironics, Inc. recalls Trilogy EVO Ventilator, EU - Model Number: EU2110X15B Trilogy Evo is intended for pediatric through adult patients we
Recalled February 26, 2020 · FDA recall Z-0945-2020 · Philips Respironics
Hazard
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439. · FDA Class I
Units
48
Sold at
US: None OUS: CH DE DK IT FR IE GB KW QA PL ES AT PH AE SI CZ
Description
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.
Product
Trilogy EVO Ventilator, EU - Model Number: EU2110X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.
Manufacturer(s)
Philips Respironics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →