RecallWatchUSA

Home / Recalls / Medical Devices

Philips Ultrasound, Inc. recalls Philips Ultrasound, Model Q-Station, with software version 3 or higher, Catalog number: 795088; Part Number: 98960538239

Recalled June 24, 2015 · FDA recall Z-1807-2015 · Philips Ultrasound

Hazard
Philips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients. · FDA Class II
Units
125 units (30 in US and 95 international)
Sold at
Worldwide Distribution - US (nationwide) and Puerto Rico; internationally to Australia, Canada, Colombia, France, Germany, Great Britain, India, Italy, Mexico, Netherlands, Norway, Pakistan, Poland, Russia, Singapore, Spain, Switzerland, Thailand, and Zimbabwe.

Description

Philips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients.

Product
Philips Ultrasound, Model Q-Station, with software version 3 or higher, Catalog number: 795088; Part Number: 989605382391 Q-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provide integrated tools that allow users to manually assess and score cardiac wall motion and export images and / or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NM, CR, MG, XA, PET, RT and X-Ray modalities for reference viewing. It support connectivity to ultrasound systems, PACS and other DICOM storage repositories.
Manufacturer(s)
Philips Ultrasound, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Philips Ultrasound, Inc. recalls Philips Ultrasound, Model Q-Station, with software version 3 or higher, Catalog number: 795088; Part Number: 98960538239 | RecallWatchUSA