Philips Ultrasound Inc recalls Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - P
Recalled May 13, 2020 · FDA recall Z-1842-2020 · Philips Ultrasound
Hazard
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury · FDA Class II
Units
N/A
Sold at
Worldwide distribution - US Nationwide distribution.
Description
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury
Product
Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies