Home / Recalls / Medical Devices
Physio-Control Inc recalls LIFEPAK 15 Monitor/Defibrillator Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a complete acute c
Recalled March 6, 2019 · FDA recall Z-0862-2019 · Physio-Control
Hazard
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and device functions. · FDA Class I
Units
13,005 units
Sold at
Worldwide Distribution
Description
Certain LIFEPAK 15 Monitors/ Defibrillators were reported to experience a lockup condition after a shock was delivered. This condition is defined as a blank monitor display with LED lights on, indicating power on the device, but no response in keypad and device functions.
Product
LIFEPAK 15 Monitor/Defibrillator Product Usage: The LIFEPAK(R) 15 Monitor/defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. The LP15 monitor/defibrillator is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 monitor/defibrillator is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
Manufacturer(s)
Physio-Control Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →