RecallWatchUSA

Home / Recalls / Medical Devices

Physio-Control, Inc. recalls LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected

Recalled January 22, 2020 · FDA recall Z-0816-2020 · Physio-Control

Hazard
Monitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button. · FDA Class II
Units
29952
Sold at
Worldwide Distribution - U.S Nationwide: GA, PA, IN, CA, KY, FL, AZ, WA, NC, MD, TX, CO, VT, MI, MA, TN, OH, CT, NY, MO, VA, IA, MN, LA, WV, MT, IL, AR, UT, NJ, OK, NV, NH, MS, NM, WI, OR, ID, SC, NE, AK, SD, AL, WY, RI, KS, ME, DE, ND, DC, HI, PR O.U.S. (Foreign): Germany, Netherlands, Hong Kong, Costa Rica, Canada, Singapore, Australia, France, Croatia, Italy , Sweden, Hungary,United Arab Emirates, Qatar, Ireland, Greece, Spain, Poland, Saudi Arabia, Reunion, Israel, United Kingdom, Mexico, South Africa, China, Switzerland, Denmark, Cayman Islands, Finland, Norway, Austria, Japan, Belgium, Czech Republic, New Zealand, Slovenia, Republic of Korea, Oman, Egypt, French Guiana, Iceland, Portugal, Latvia, Lebanon, Chile, Kuwait, Bosnia and Herzegovina, Yemen, Eritrea, Martinique, New Caledonia, Kenya, Romania, Guinea, Luxembourg, Kazakhstan, Papua New Guinea, Brazil, Venezuela, Bolivarian Republic of, Trinidad and Tobago, Algeria, Turkey, Ukraine, Bahrain, Congo, the Democratic Republic of the, Nigeria, Lao People's Democratic Republic, Russian Federation, Cyprus, Pakistan, Faroe Islands, Gabon,Libya, Thailand, Indonesia, Lithuania, Georgia, Malta, Macedonia, the former Yugoslav Republic of, Suriname, San Marino, Morocco, Guam, Taiwan, Province of China, Malaysia,Serbia, Brunei Darussalam, Argentina, Virgin Islands, U.S., Barbados, Mali, Iraq, Azerbaijan, Botswana, Kosovo, Cambodia, Cura¿ao, Ecuador, Aruba

Description

Monitor/Defibrillator may not deliver a shock after the "Shock" button on the keypad is pressed as a result of oxidation that has formed over time within the button.

Product
LIFEPAK 15 MONITOR/DEFIBRILLATOR that was either manufactured with or received an upgrade kit that contained an affected keypad.
Manufacturer(s)
Physio-Control, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov