RecallWatchUSA

Home / Recalls / Medical Devices

Polymer Technology Systems, Inc. recalls CHOL+HDL+GLU test strips, REF 2412

Recalled May 29, 2019 · FDA recall Z-1373-2019 · Polymer Technology Systems

Hazard
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed · FDA Class II
Units
67643 total
Sold at
Domestic distribution nationwide.

Description

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed

Product
CHOL+HDL+GLU test strips, REF 2412
Manufacturer(s)
Polymer Technology Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Polymer Technology Systems, Inc. recalls CHOL+HDL+GLU test strips, REF 2412 | RecallWatchUSA