Polymer Technology Systems, Inc. recalls Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036

Recalled January 29, 2020 · FDA recall Z-0834-2020 · Polymer Technology Systems

Hazard
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating. · FDA Class II
Units
44,448
Sold at
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

Description

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Product
Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036
Manufacturer(s)
Polymer Technology Systems, Inc.