RecallWatchUSA

Home / Recalls / Medical Devices

Polymer Technology Systems, Inc. recalls Lipid Panel test strips, REF 1710, also private labeled under Henry Schein as REF 570-0414

Recalled May 29, 2019 · FDA recall Z-1371-2019 · Polymer Technology Systems

Hazard
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed. · FDA Class II
Units
67643 total
Sold at
Domestic distribution nationwide.

Description

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed.

Product
Lipid Panel test strips, REF 1710, also private labeled under Henry Schein as REF 570-0414
Manufacturer(s)
Polymer Technology Systems, Inc.