RecallWatchUSA

Home / Recalls / Medical Devices

Polymer Technology Systems, Inc. recalls PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Num

Recalled January 29, 2020 · FDA recall Z-0833-2020 · Polymer Technology Systems

Hazard
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating. · FDA Class II
Units
44,448 total
Sold at
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

Description

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Product
PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Numbers 2702
Manufacturer(s)
Polymer Technology Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Polymer Technology Systems, Inc. recalls PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Num | RecallWatchUSA