Precision Medical, Inc. recalls Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 l
Recalled November 25, 2015 · FDA recall Z-0272-2016 · Precision Medical
Hazard
These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25). · FDA Class II
Units
315
Sold at
Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.
Description
These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).
Product
Easy Dial Oxygen Regulator, 168708D and 168715D. Eight liter from .0.25 to 8 liters, 15 liter from 0.5 liter to 15 liters of oxygen per minute.
Manufacturer(s)
Precision Medical, Inc.