RecallWatchUSA

Home / Recalls / Medical Devices

Primus Medical LLC recalls AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, O

Recalled November 14, 2012 · FDA recall Z-0270-2013 · Primus Medical

Hazard
The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse. · FDA Class II
Units
342 Beds
Sold at
Nationwide distribution: USA including states of: AZ, IL, KY, MD, MI, OH, OR, PA, TX and VA.

Description

The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.

Product
AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
Manufacturer(s)
Primus Medical LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov