RecallWatchUSA

Home / Recalls / Medical Devices

PROCEPT BIOROBOTICS CORPORATION recalls AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distri

Recalled April 17, 2019 · FDA recall Z-1110-2019 · PROCEPT BIOROBOTICS

Hazard
Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure. · FDA Class II
Units
214
Sold at
Worldwide Distribution: US (nationwide) state to: TX, FL, IN, CA, IL, VA, MI, PA.; and (OUS) countries of: Italy, Germany, France, Poland, United Kingdom, Spain, Canada and Australia.

Description

Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.

Product
AquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.S. Device Description The AQUABEAM Robotic System is intended for use in patients suffering from lower urinary tract symptoms resulting from benign prostatic hyperplasia (BPH).
Manufacturer(s)
PROCEPT BIOROBOTICS CORPORATION

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov