PTW-FREIBURG recalls BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is perf

Recalled September 20, 2023 · FDA recall Z-2577-2023 · PTW-FREIBURG

Hazard
Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file. · FDA Class II
Units
42 US
Sold at
US Nationwide distribution.

Description

Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.

Product
BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360)
Manufacturer(s)
PTW-FREIBURG
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov