RecallWatchUSA

Home / Recalls / Medical Devices

Qiagen Sciences LLC recalls QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223

Recalled January 12, 2022 · FDA recall Z-0455-2022 · Qiagen Sciences

Hazard
The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur. · FDA Class II
Units
376 kits
Sold at
Domestic distribution to AL, DC, KY, MD, ME, NJ, OK< PA, and TX. No foreign distribution.

Description

The firm identified sixteen (16) faulty cartridges in LOT 210209, which contain incorrect primer-probe mixes in the reaction chambers. If an affected cartridge is used, false-negative, false-positive, or both false-negative and false-positive results could occur.

Product
QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223
Manufacturer(s)
Qiagen Sciences LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Qiagen Sciences LLC recalls QIAstat-Dx Respiratory SARS-CoV-2 Panel (US IVD under FDA EUA 200075); REF 691223 | RecallWatchUSA