RecallWatchUSA

Home / Recalls / Medical Devices

QiG Group LLC recalls Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2412 ( 2 x 12 channel)

Recalled March 30, 2016 · FDA recall Z-1193-2016 · QiG Group

Hazard
The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator. · FDA Class II
Units
34
Sold at
Distributed in Germany and Luxembourg.

Description

The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.

Product
Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2412 ( 2 x 12 channel)
Manufacturer(s)
QiG Group LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

QiG Group LLC recalls Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2412 ( 2 x 12 channel) | RecallWatchUSA