RecallWatchUSA

Home / Recalls / Medical Devices

Qualigen Inc recalls FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP

Recalled October 16, 2024 · FDA recall Z-0022-2025 · Qualigen

Hazard
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected. · FDA Class II
Units
60
Sold at
US: TX, MS, NY, OH, MT, WA, PA, MN, SC, GA, NC, NV, WY, AK, SD, KS, NE, VT, AL, MA, OK, UT, WI, IA, WV. OUS: Germany, Switzerland

Description

Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.

Product
FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer
Manufacturer(s)
Qualigen Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Qualigen Inc recalls FastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP | RecallWatchUSA