RecallWatchUSA

Home / Recalls / Medical Devices

Quest Medical, Inc. recalls Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The d

Recalled June 3, 2020 · FDA recall Z-2107-2020 · Quest Medical

Hazard
Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention. · FDA Class II
Units
5 units
Sold at
US Nationwide distribution including in the states of NY, PA, TN, WV. OUS: None

Description

Additive and Arrest Agent Labels on the MPS Top Cover Replacement Kit, a replacement for worn or damaged MPS Console Top Covers for cardiopulmonary bypass heat exchanger, used to deliver whole blood to the heart during open heart surgery, were swapped and incorrectly placed. This could lead to over delivery of potassium by the user requiring medical intervention.

Product
Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The device is a component of the MPS Console used as replacements for worn or damaged MPS Console top covers)
Manufacturer(s)
Quest Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Quest Medical, Inc. recalls Ref 801168, MPS Top Cover Replacement Kit, 1 UNITS, UDI: (01)2 0634624 81168 7 (11)190916 (17)290913 (10)059135 (The d | RecallWatchUSA