Quidel Corporation recalls Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immun

Recalled April 4, 2018 · FDA recall Z-1183-2018 · Quidel

Hazard
The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported. · FDA Class II
Units
482 units
Sold at
US Distribution

Description

The calibration for the affected batch of analyzers was incorrectly set during the manufacturing process which may lead to a failure when calibration is first performed. There may also be e remote possibility that incorrect qualitative test results for patient samples near the assay cutoff may be reported.

Product
Sofia 2 Analyzer Catalog Number 20299 Sofia 2 is a bench top analyzer intended to be used with Cassette-based immunofluorescent in vitro diagnostic assays . Sofia 2 is intended for professional and laboratory use.
Manufacturer(s)
Quidel Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov