RecallWatchUSA

Home / Recalls / Medical Devices

Ra Medical Systems, Inc. recalls DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disea

Recalled December 23, 2020 · FDA recall Z-0569-2021 · Ra Medical Systems

Hazard
Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase. · FDA Class II
Units
36 devices
Sold at
US Distribution to states of: AL, AZ, CA, FL, IL, IN, KY, LA, MD, NC, NV, NY, OK, OR, TX, and WA; and to OUS country: Japan.

Description

Due to footswitch not meeting required specification for protection from ingress of solids or liquids. If spilling of liquids near or on the footswitch, it may fail in the 'on' position and allow the laser to continue to lase.

Product
DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disease. Indications: For crossing chronic total occlusions in patients with symptomatic infrainguinal lower extremity vascular disease.
Manufacturer(s)
Ra Medical Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Ra Medical Systems, Inc. recalls DABRA Laser (RA-308 Excimer Laser) Intended Use: For use in ablating a channel in occlusive peripheral vascular disea | RecallWatchUSA