Radiometer America Inc recalls AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality contro
Recalled October 11, 2017 · FDA recall Z-3237-2017 · Radiometer America
Hazard
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg) · FDA Class II
Units
1669 Boxes
Sold at
Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.
Description
Radiometer has recently become aware that a few of the AUTOCHECK 5+ level 2 ampoules, will give incorrect results for pCO2 and pO2. A defective QC ampoule will for: " pCO2 give a result close to or just outside of the low end of the control range. (4.92 5.85 kPa or 36.9 46.9 mmHg) " pO2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kPa (104-107 mmHg)
Product
AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers
Manufacturer(s)
Radiometer America Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →