Radiometer Medical ApS recalls ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended fo
Recalled June 3, 2020 · FDA recall Z-2139-2020 · Radiometer Medical ApS
Hazard
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors. · FDA Class II
Units
1233 units
Sold at
Worldwide distribution - US Nationwide. Also distributed OUS.
Description
The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
Product
ABL90 FLEX, model no. 393-090; and Spare Part Hemolyzer Units (902-980): R0526 - R0941 - Product Usage: are intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point-of-care setting.
Manufacturer(s)
Radiometer Medical ApS
Made in
Denmark