Home / Recalls / Medical Devices
Raye's Inc. recalls SW Bari Lift & Transfer, Model Number 38060000
Recalled April 24, 2024 · FDA recall Z-1595-2024 · Raye's
Hazard
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device. · FDA Class II
Units
1,149 units
Sold at
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Description
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
Product
SW Bari Lift & Transfer, Model Number 38060000
Manufacturer(s)
Raye's Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →