RecallWatchUSA

Home / Recalls / Medical Devices

RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RaySta

Recalled January 28, 2026 · FDA recall Z-1110-2026 · RAYSEARCH LABORATORIES AB

Hazard
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended. · FDA Class II
Units
0
Sold at
Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Description

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RayStation v2025, RayStation v2025 SP1 Product Description: Radiation Therapy Treatment Planning System Component: No
Manufacturer(s)
RAYSEARCH LABORATORIES AB
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

RAYSEARCH LABORATORIES AB recalls Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version: RaySta | RecallWatchUSA