RAYSEARCH LABORATORIES AB recalls RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. T

Recalled January 24, 2018 · FDA recall Z-0380-2018 · RAYSEARCH LABORATORIES AB

Hazard
For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient. · FDA Class II
Units
117 units
Sold at
US Nationwide Distribution - FL, WA, AZ, TN, MD, TX, FL, MI, Washington D.C.

Description

For the proton QA preparation module in RayStation 6 (including SP1 and SP2), if the snout position or gap is modified in the QA module, the dose in the QA module may be computed for a different setup than what is used for QA measurements. This could result in an incorrect dosage being provided to a patient.

Product
RayStation Product Usage: is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.