RAYSEARCH LABORATORIES AB recalls RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of r
Recalled June 25, 2014 · FDA recall Z-1806-2014 · RAYSEARCH LABORATORIES AB
Hazard
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution. · FDA Class II
Units
25 sites/245 units (US), 25 sites/124 units (foreign)
Sold at
Worldwide Distribution-USA (nationwide) including the states of AZ, CA, FL, IL, MA, MD, MO, MS, NJ, NV, NY, OH, PA, TN, VA, and WA, and the countries of Australia, Belgium, Germany, France, Italy, New Zealand, Republic of Korea, Switzerland, Spain, and United Kingdom.
Description
This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of an ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.
Product
RayStation software version 4.0.3.4 RayStation is a software system designed for treatment planning and analysis of radiation therapy.
Manufacturer(s)
RAYSEARCH LABORATORIES AB
Made in
Sweden
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →