RecallWatchUSA

Home / Recalls / Medical Devices

RAYSEARCH LABORATORIES AB recalls RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.

Recalled December 2, 2020 · FDA recall Z-0478-2021 · RAYSEARCH LABORATORIES AB

Hazard
If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI. · FDA Class II
Units
2768
Sold at
US Nationwide distribution.

Description

If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.

Product
RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user needs.
Manufacturer(s)
RAYSEARCH LABORATORIES AB
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

RAYSEARCH LABORATORIES AB recalls RayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8. | RecallWatchUSA