Home / Recalls / Medical Devices
RAYSEARCH LABORATORIES AB recalls Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10
Recalled March 29, 2023 · FDA recall Z-1281-2023 · RAYSEARCH LABORATORIES AB
Hazard
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle. · FDA Class II
Units
10770 systems
Sold at
US Nationwide
Description
It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.
Product
Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0. Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs
Manufacturer(s)
RAYSEARCH LABORATORIES AB
Made in
Sweden
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →