RecallWatchUSA

Home / Recalls / Medical Devices

Reliance Medical Products Inc recalls Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of syst

Recalled April 18, 2018 · FDA recall Z-1318-2018 · Reliance Medical Products

Hazard
Product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing · FDA Class II
Units
933 ( of which 97 has been quarantined)
Sold at
Worldwide Distribution - USA(nationwide) Distribution to the states of : NJ, CA, OH, VA, MA, MN, IL , UT, TN, AR, LA, AZ, PA, TX, IA and HI., and to the countries of : Costa Rica, Jamaica, South Africa, New Zealand, South Korea, Dubai, Saudi Arabia and Jordan.

Description

Product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing

Product
Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position ophthalmological examination and diagnostic equipment, is to store and position for examination a device commonly known as a refractor.
Manufacturer(s)
Reliance Medical Products Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Reliance Medical Products Inc recalls Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of syst | RecallWatchUSA